Guide 02

Calibration steps

A repeatable path from the first check of the gauge to the moment the record sheet is closed.

Before the reading

  1. Identify the gauge. Match the tag in the register with the gauge on the bench.
  2. Check the reference device. Note its label and the unit used for the check.
  3. Note the setting. Record temperature, humidity, and any condition that could affect the reading.
  4. Prepare the gauge. Clean the contact area and let the instrument settle.

Zero check

Set the gauge to its zero point and observe the result. If the rest position is not at zero, write the value down before making any adjustment.

Reference points

Work through the planned points in order. At each point, enter the set value, the observed value, and the unit. Sternwick places the deviation beside them using the same numeric style.

Sample reference points for a pressure gauge
Point Set Observed Deviation
14.00 bar4.01 bar+0.01 bar
28.00 bar7.99 bar-0.01 bar
312.00 bar12.02 bar+0.02 bar

Repeat readings

Where the method asks for a repeat, enter it as its own line. Keeping repeats separate makes the path easier to follow than replacing the first value.

Close the record

Review each deviation, add a short closing note, and confirm the result. The sheet then gathers the gauge identity, setting, steps, readings, and result in one view. For the field list, see certificate fields.

Clear wording: use plain notes such as “settled for two minutes” or “reading stable at third point.” Short observations travel well with the record.